The YL9300 HPLC is an integrated HPLC system in a compact design including vacuum degasser, quaternary pump with a mixing valve and UV/Vis detector at a competitive price.

Its newly innovated focus lens in a monochromator reduces mobile-phase RI effect and provides enhanced baseline stability with a substantially minimized noise level, especially in short-wavelength range, which decreases solvent related peaks.

  • Benefits
  • Specifications
  • Demo
  • Quote
  • Downloads

Benefits

Compact Design
  • Integrated LPG (Low-Pressure Gradient) HPLC system: Vacuum Degasser + Quaternary Pump + UV Detector
  • Simple maintenance: Pump & Vacuum Degasser modules are located in a sliding drawer

Perfect Performance
  • Increased mobile phase composition efficiency
  • Reduced solvent peak influence
  • Narrow spectral bandwidth
  • Enhanced baseline stability with minimized noise level

Cost-Effective
  • Low-cost maintenance
  • Competitively priced

Specifications

Vacuum Degasser

  • Number of channels: 4 CHs
  • Maximum flow rate: 10 ml/min per channel
  • Internal volume per channel: 925 ul per channel
  • Solvent contact materials: Teflon AF, PEEK, and Glass-filled PTFE


Quaternary Pump

  • Flow range: 0.001-10 ml/min
  • Flow rate accuracy: ≦ ±1% at 1 ml/min
  • Flow rate precision: < 0.1% RSD at 1 ml/min
  • Number of eluent lines: 4
  • Pressure pulsation: ≦ ±0.5 % at 1 ml/min


UV/VIS Detector

  • Wavelength range: 190~600 nm (Standard) /190~900 nm (Option)
  • Spectral bandwidth: 5 nm
  • Wavelength accuracy: ±1 nm
  • Wavelength precision: ±0.1 nm
  • Linearity: > 99.5% at 2.5 AU (Acetone, 254 nm)
  • Noise level: < ±0.35 X 10-5 AU, 254 nm, dry cell
  • Drift: < 1 X 10-4 AU/hour


System Information

  • Data communications: LAN
  • Dimension: 375 X 470 X 545 mm (W X H X D)
  • Weight: 27 Kg
  • Safety & maintenance: Leak detection, Diagnostics, Error detection
  • Power consumption: 150 W


YL9330 Column Compartment

  • Temperature range : 4°C (Cooling) - 90 °C
  • Temperature stability : ±0.05°C
  • Temperature accuracy : ±0.5°C, with 2-point temp. calibration
  • Temperature programs : 40 Steps
  • Column capacity : three columns up to 300mm length (max OD: up to 18mm)
  • Column switching (Option): automatically 6-port SS(PEEK) valve up to 2 ea
  • Communications: LAN
  • Safety & maintenance : Leak detection, Diagnostics, Error detection
  • Dimensions : 185 x 476 x 480mm (W x H x D)
  • Line Voltage : 100-240VAC, ±10%, automatic voltage selection
  • Line frequency : 50/60Hz, ±5%
  • Power consumption : 150W

Demo

 

Quote

 

Quote